天天摸天天碰天天添中文字幕,国产区点击进入在线影院,无码人妻一区二区免费AV,亚洲精品国产av现线

產(chǎn)品中心

一步法人血管內(nèi)皮生長因子165(VEGF165)酶聯(lián)免疫吸附檢測試劑盒
ELK027ES
規(guī)格: 價(jià)格:
64T ¥2240.00
112T ¥3200.00

Overview

Product name: EasyStep Human VEGF165(Vascular Endothelial Growth Factor 165) ELISA Kit
Reactivity: Human
Alternative Names: VEGF165
Assay Type: Sandwich
Sensitivity: 15 pg/mL
Standard: 2000 pg/mL
Detection Range: 31.25-2000 pg/mL
Sample Type: Serum, plasma
Assay Length: 1.5h
Research Area: Cytokine;Tumor immunity;
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human VEGF165, and the Human VEGF165 standard plate wells that pre-coated using protein-related techniques are provided separately. Standard/Sample Diluent Buffer or samples are added to the appropriate microtiter plate wells ,then added a HRP-conjugated antibody specific to Human VEGF165. After TMB substrate solution is added, only those wells that contain Human VEGF165 and HRP-conjugated antibody will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human VEGF165 in the samples is then determined by comparing the OD of the samples to the standard curve.

標(biāo)準(zhǔn)曲線

Concentration (pg/mL) OD Corrected OD
2000.00 2.126 2.033
1000.00 1.731 1.638
500.00 1.114 1.021
250.00 0.855 0.762
125.00 0.532 0.439
62.50 0.335 0.242
31.25 0.226 0.133
0.00 0.093 0.000

精密度

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

回收率

Matrices listed below were spiked with certain level of recombinant VEGF165 and the recovery rates were calculated by comparing the measured value to the expected amount of VEGF165 in samples.
Matrix Recovery range Average
serum(n=5) 95-108% 101%
EDTA plasma(n=5) 83-97% 90%
Heparin plasma(n=5) 80-97% 88%

線性

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of VEGF165 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 85-97% 87-96% 85-96% 85-95%
EDTA plasma(n=5) 95-106% 87-96% 82-90% 92-101%
Heparin plasma(n=5) 85-92% 95-103% 97-105% 93-102%

產(chǎn)品成分表

關(guān)閉

在線咨詢

Online consultation

  • 在線咨詢
  • 技術(shù)支持

關(guān)注微信公眾號

微信掃一掃立即咨詢

微信掃一掃立即咨詢

關(guān)閉
鄂伦春自治旗| 神农架林区| 天门市| 平利县| 东至县| 清水县| 滕州市| 虎林市| 临海市| 故城县| 琼中| 宁远县| 正蓝旗| 麻江县| 繁昌县| 靖远县| 钟山县| 晋城| 雅江县| 温州市| 沙雅县| 枝江市| 巍山| 随州市| 阿合奇县| 平顶山市| 游戏| 福鼎市| 麟游县| 仙居县| 沅江市| 新竹市| 长岭县| 虞城县| 南充市| 西青区| 新民市| 常山县| 沂水县| 栾川县| 宜章县|