天天摸天天碰天天添中文字幕,国产区点击进入在线影院,无码人妻一区二区免费AV,亚洲精品国产av现线

產(chǎn)品中心

一步法人促卵泡素(FSH)酶聯(lián)免疫吸附檢測試劑盒
ELK031ES
規(guī)格: 價格:
64T ¥2240.00
112T ¥3200.00

Overview

Product name: EasyStep Human FSH(Follicle Stimulating Hormone) ELISA Kit
Reactivity: Human
Alternative Names: FSH-B; FSH-beta
Assay Type: Sandwich
Sensitivity: 0.016 mIU/mL
Standard: 100 mIU/mL
Detection Range: 1.56-100 mIU/mL
Sample Type: Serum, plasma
Assay Length: 1.5h
Research Area: Endocrinology;Reproductive science;Genetic science;Hormone metabolism;
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human FSH, and the Human FSH standard plate wells that pre-coated using protein-related techniques are provided separately. Standard/Sample Diluent Buffer or samples are added to the appropriate microtiter plate wells ,then added a HRP-conjugated antibody specific to Human FSH. After TMB substrate solution is added, only those wells that contain Human FSH and HRP-conjugated antibody will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human FSH in the samples is then determined by comparing the OD of the samples to the standard curve.

標準曲線

Concentration (mIU/mL) OD Corrected OD
100.00 1.972 1.886
50.00 1.554 1.468
25.00 1.185 1.099
12.50 0.831 0.745
6.25 0.567 0.481
3.13 0.386 0.300
1.57 0.189 0.103
0.00 0.086 0.000

精密度

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

回收率

Matrices listed below were spiked with certain level of recombinant FSH and the recovery rates were calculated by comparing the measured value to the expected amount of FSH in samples.
Matrix Recovery range Average
serum(n=5) 82-95% 88%
EDTA plasma(n=5) 78-96% 87%
Heparin plasma(n=5) 95-107% 101%

線性

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of FSH and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 91-99% 83-89% 88-104% 97-103%
EDTA plasma(n=5) 85-99% 82-96% 92-101% 82-90%
Heparin plasma(n=5) 79-92% 80-93% 81-98% 89-102%

產(chǎn)品成分表

關閉

在線咨詢

Online consultation

  • 在線咨詢
  • 技術(shù)支持

關注微信公眾號

微信掃一掃立即咨詢

微信掃一掃立即咨詢

關閉
西华县| 平山县| 隆昌县| 石渠县| 海丰县| 洪湖市| 肥城市| 鄂尔多斯市| 锦州市| 桃园市| 泸州市| 乐安县| 讷河市| 平罗县| 惠州市| 鄂托克旗| 中西区| 东源县| 太谷县| 红原县| 屯留县| 丰都县| 额尔古纳市| 拉孜县| 桐柏县| 南郑县| 六安市| 得荣县| 张北县| 佛冈县| 竹北市| 建瓯市| 东莞市| 日喀则市| 大安市| 开封市| 云龙县| 屏边| 浮梁县| 东安县| 桐梓县|